ASTM D5229 · Moisture uptake calibration
Characterizes moisture uptake and conditioning of composites before comparing properties at a defined moisture state.
CDS sign-in and suitable licensed access may be required.
Recorded instructional case · Workbench captures and saved calculation data, October 2026. This website preview does not run the test.
Workbench conditioning setup

Uptake & calibration
Mass gain (%)
Time (h)
- Homogeneous slab uptake
- Entered early measurements
View plotted values
| Series | Time (h) | Mass gain (%) |
|---|---|---|
| Homogeneous slab uptake | 0 | 0 |
| Homogeneous slab uptake | 1 | 0.05000000000000001 |
| Homogeneous slab uptake | 2 | 0.07071067811865477 |
| Homogeneous slab uptake | 3 | 0.08660254037844388 |
| Homogeneous slab uptake | 4 | 0.10000000000000002 |
| Homogeneous slab uptake | 5 | 0.11180339887498951 |
| Homogeneous slab uptake | 6 | 0.12247448713915891 |
| Homogeneous slab uptake | 7 | 0.13228756555322954 |
| Homogeneous slab uptake | 8 | 0.14142135623730953 |
| Homogeneous slab uptake | 9 | 0.15000000000000002 |
| Homogeneous slab uptake | 10 | 0.158113883008419 |
| Homogeneous slab uptake | 11 | 0.16583123951777 |
| Homogeneous slab uptake | 12 | 0.17320508075688776 |
| Homogeneous slab uptake | 13 | 0.18027756377319948 |
| Homogeneous slab uptake | 14 | 0.18708286933869708 |
| Homogeneous slab uptake | 15 | 0.19364916731037085 |
| Homogeneous slab uptake | 16 | 0.20000000000000004 |
| Homogeneous slab uptake | 17 | 0.20615528128088306 |
| Homogeneous slab uptake | 18 | 0.21213203435596428 |
| Homogeneous slab uptake | 19 | 0.21794494717703372 |
| Homogeneous slab uptake | 20 | 0.22360679774997902 |
| Homogeneous slab uptake | 21 | 0.2291287847475303 |
| Homogeneous slab uptake | 22 | 0.23452078799032894 |
| Homogeneous slab uptake | 23 | 0.23979157616317848 |
| Homogeneous slab uptake | 24 | 0.244948974271744 |
| Homogeneous slab uptake | 25 | 0.24999999998370925 |
| Homogeneous slab uptake | 26 | 0.25495097564191005 |
| Homogeneous slab uptake | 27 | 0.2598076210530603 |
| Homogeneous slab uptake | 28 | 0.26457513093646867 |
| Homogeneous slab uptake | 29 | 0.26925824002206955 |
| Homogeneous slab uptake | 30 | 0.2738612781218541 |
| Homogeneous slab uptake | 31 | 0.27838821699871097 |
| Homogeneous slab uptake | 32 | 0.2828427104768154 |
| Homogeneous slab uptake | 33 | 0.28722812894629424 |
| Homogeneous slab uptake | 34 | 0.2915475891893309 |
| Homogeneous slab uptake | 35 | 0.2958039802787834 |
| Homogeneous slab uptake | 36 | 0.29999998616183754 |
| Homogeneous slab uptake | 37 | 0.3041381054313831 |
| Homogeneous slab uptake | 38 | 0.3082206687001319 |
| Homogeneous slab uptake | 39 | 0.31224985392204585 |
| Homogeneous slab uptake | 40 | 0.31622769994868216 |
| Homogeneous slab uptake | 41 | 0.3201561185617242 |
| Homogeneous slab uptake | 42 | 0.3240369051850711 |
| Homogeneous slab uptake | 43 | 0.32787174844868505 |
| Homogeneous slab uptake | 44 | 0.33166223875062006 |
| Homogeneous slab uptake | 45 | 0.33540987594225247 |
| Homogeneous slab uptake | 46 | 0.33911607624385853 |
| Homogeneous slab uptake | 47 | 0.3427821784826929 |
| Homogeneous slab uptake | 48 | 0.3464094497331135 |
| Homogeneous slab uptake | 49 | 0.3499990904276069 |
| Homogeneous slab uptake | 50 | 0.3535522389985254 |
| Homogeneous slab uptake | 51 | 0.3570699761026106 |
| Homogeneous slab uptake | 52 | 0.3605533284737842 |
| Homogeneous slab uptake | 53 | 0.3640032724440132 |
| Homogeneous slab uptake | 54 | 0.3674207371671856 |
| Homogeneous slab uptake | 55 | 0.37080660757671435 |
| Homogeneous slab uptake | 56 | 0.3741617271039488 |
| Homogeneous slab uptake | 57 | 0.3774869001812987 |
| Homogeneous slab uptake | 58 | 0.38078289455123004 |
| Homogeneous slab uptake | 59 | 0.3840504433998839 |
| Homogeneous slab uptake | 60 | 0.38729024733196804 |
| Homogeneous slab uptake | 61 | 0.3905029762017349 |
| Homogeneous slab uptake | 62 | 0.3936892708132448 |
| Homogeneous slab uptake | 63 | 0.39684974450168564 |
| Homogeneous slab uptake | 64 | 0.39998498460628085 |
| Homogeneous slab uptake | 65 | 0.40309555384420626 |
| Homogeneous slab uptake | 66 | 0.4061819915939635 |
| Homogeneous slab uptake | 67 | 0.40924481509579735 |
| Homogeneous slab uptake | 68 | 0.41228452057597753 |
| Homogeneous slab uptake | 69 | 0.4153015843010903 |
| Homogeneous slab uptake | 70 | 0.4182964635678782 |
| Homogeneous slab uptake | 71 | 0.4212695976336396 |
| Homogeneous slab uptake | 72 | 0.42422140859170787 |
| Homogeneous slab uptake | 73 | 0.42715230219612443 |
| Homogeneous slab uptake | 74 | 0.4300626686392186 |
| Homogeneous slab uptake | 75 | 0.43295288328548853 |
| Homogeneous slab uptake | 76 | 0.43582330736485464 |
| Homogeneous slab uptake | 77 | 0.4386742886281024 |
| Homogeneous slab uptake | 78 | 0.4415061619670728 |
| Homogeneous slab uptake | 79 | 0.44431925000195016 |
| Homogeneous slab uptake | 80 | 0.44711386363779226 |
| Homogeneous slab uptake | 81 | 0.4498903025922788 |
| Homogeneous slab uptake | 82 | 0.45264885589648274 |
| Homogeneous slab uptake | 83 | 0.4553898023703392 |
| Homogeneous slab uptake | 84 | 0.458113411074341 |
| Homogeneous slab uptake | 85 | 0.4608199417388861 |
| Homogeneous slab uptake | 86 | 0.46350964517258286 |
| Homogeneous slab uptake | 87 | 0.4661827636507333 |
| Homogeneous slab uptake | 88 | 0.46883953128511635 |
| Homogeneous slab uptake | 89 | 0.47148017437611967 |
| Homogeneous slab uptake | 90 | 0.4741049117481926 |
| Homogeneous slab uptake | 91 | 0.4767139550695286 |
| Homogeneous slab uptake | 92 | 0.47930750915681797 |
| Homogeneous slab uptake | 93 | 0.4818857722658667 |
| Homogeneous slab uptake | 94 | 0.4844489363688148 |
| Homogeneous slab uptake | 95 | 0.4869971874186487 |
| Homogeneous slab uptake | 96 | 0.4895307056016539 |
| Homogeneous slab uptake | 97 | 0.49204966557841867 |
| Homogeneous slab uptake | 98 | 0.4945542367139578 |
| Homogeneous slab uptake | 99 | 0.49704458329749746 |
| Homogeneous slab uptake | 100 | 0.4995208647524231 |
| Entered early measurements | 1 | 0.05 |
| Entered early measurements | 4 | 0.1 |
Set up in CDS
- Link the slab thickness and equilibrium moisture gain.
- Enter two early mass-gain measurements and their times.
- Fit diffusivity from the square-root-time slope.
- Compare entered measurements with the uptake curve.
Inputs for this recorded case
- 2 mm linked slab
- 16 plies / 2 mm
- Entered uptake calibration
- Explicit source calibration/properties
These are teaching-case inputs, not ASTM-prescribed specimen dimensions or acceptance criteria.
Readout & interpretation
- Fitted diffusivity
- 5.454e-7 mm²/s
- Equilibrium mass gain
- 1 %
- Fit intercept
- 0 %
- Final predicted mass gain
- 0.4995 %
Initially dry homogeneous slab with both faces exposed. Edge ingress is excluded.
About this ASTM reference
Standard Test Method for Moisture Absorption Properties and Equilibrium Conditioning of Polymer Matrix Composite Materials
Active catalog edition checked October 10, 2026. The official standard defines the physical method; the CDS example below has a narrower modeling scope. Full procedures remain on ASTM’s site.
Find this setup in Workbench
Opens the connected guided study. Select the matching ASTM setup in the Test menu to inspect its fixture.
What this CDS case represents
Read the actual ASTM standard ↗Initially dry homogeneous slab with both faces exposed. Edge ingress is excluded.
This conditioning calibration provides its own curves and metrics. Synchronized failure playback is not supported; a mounted fixture view does not imply that contact, crack growth or impact history is solved.
Model basis · equation and interpretation
D = π [t × slope/(4M∞)]²; M(T)/M∞ = 1 − (8/π²) Σ exp[−(2j+1)²π²DT/t²]/(2j+1)².
t is total slab thickness, T elapsed time, M mass gain, M∞ equilibrium mass gain and j a nonnegative integer in the series. Two increasing early measurements define slope = ΔM/Δ√T; both must be at or below 50% of equilibrium. The initially dry homogeneous slab has both faces exposed and constant diffusivity D (mm²/s). No edge ingress, temperature shifting or strength degradation is inferred. Saved material records are not changed.
Instructional model comparison, not experimental validation or ASTM certification. Physical testing requires the licensed standard and a qualified test program.